Bentel Ch Questions

Terms

undefined, object
copy deck
Year in which Roentgen was internationally accepted as unit for x-rays
1928
Year rad was recommended by ICRU as a unit of absorbed dose
1953
Primary reason betatrons were unpopular for use in rad therapy
Low dose rate and limited field size
Radiation therapy units operating from 50 - 120 kVp
superficial units
Orthovoltage x-ray units operate in this range of energy
150 - 500 kVp
SSD for othovoltage units
50 - 70 cm
HVT for superficial units.
1 - 8 mm Al
Treatment distances for superficial units
15 - 20 cm
The dose rate is high for orthovoltage units for this reason.
Short treatment distances
Depth to maximum dose for orthovoltage machines and depth to 90%
Close to skin surface, depth to 90% is 2 cm
the dose monitoring system in the beam path of a linac monitors these three things.
Dose rate, integrated dose, beam symmetry
Year in which Co-60 unit invented
1951
What determines the sharpness of the beam edges in a linac?
secondary collimator
How does the depth to maximum dose depend on field size, SSD, and oblique incidence?
Decreases with increasing field size and oblique incidence, is independent of SSD
neuroblastoma origination
sympathetic nervous system
Define the abdomen
lies between thorax and pelvis, contains peritoneal cavity
dose under a 5-HVT kidney block vs. middepth on central axis is...
10-20% depending on energy and FS